Microbiology FDA 483 (Sentara Enterprises): The movement of first air in the ISO 5 area is blocked or disrupted

Observation

Personnel were observed conducting aseptic manipulations where the movement of “first air” in the ISO 5 area is blocked or disrupted

Facility Type

Sterile Manufacturing

Firm / Company

Sentara Enterprises

System

Quality System, Laboratory Controls, Facility and Equipment System

Details

Specifically, on July 11, 2023 while Technician *** was working in ISO 5 classified *** Laminar airflow hood (equipment ID ***), producing MYCAMINE 100MG IN 100ML HOMEPUMP Rx *** was observed blocking first pass air with left hand (holding small vial) while making aseptic connections.

Record Date

03/2024

FEI Number

3025984445

Country

Virginia

Citation Type

483

Audit Duration (Days)

6

Investigators

Tekalign Wondimu, Sena G Dissmeyer
FDA 483 relevant to Microbiology

Disclaimer: The purpose of this article is to provide helpful information on FDA 483s related to microbiological issues. This article represents 483 information in Tabular form for quick access and easy navigation. Information for this article is derived from “Form FDA 483” (Inspectional Observation). This article is not intended to replace the information provided by the FDA, nor is it guaranteed to be complete, always refer official FDA site for details.