Microbiology FDA 483 (Eugia SEZ): Equipment is not suitably located to facilitate operations for its intended use
Observation
Facility Type
Firm / Company
Eugia SEZ Private Limited
System
Details
Specifically,
Your film’s No. 04 Continuous Nonviable Particle Counter (NVPC), located near the *** conveyor area of *** Line *** is not located in the position depicted in **** Filling Machine Lay-out NVPC”. Document No. APL-XVI-PR-EQP-002-02. It is actually located inside the Grade A area, near the *** accessible through *** No *** of RABS **, which is separated from the conveyor area by an ***. As such, there is no NVPC data to represent conditions during *** interventions that require movement of *** conveyor, or of any interventions performed through *** RABS *** No ***.
Record Date
FEI Number
Country
Citation Type
Audit Duration (Days)
Investigators
More observations from same firm / company
FDA 483 relevant to Microbiology
Disclaimer: The purpose of this article is to provide helpful information on FDA 483s related to microbiological issues. This article represents 483 information in Tabular form for quick access and easy navigation. Information for this article is derived from “Form FDA 483” (Inspectional Observation). This article is not intended to replace the information provided by the FDA, nor is it guaranteed to be complete, always refer official FDA site for details.