Microbiology FDA 483 (Zydus Lifesciences): Procedures to prevent microbiological contamination did not include adequate validation of the aseptic process
Observation
Facility Type
Firm / Company
Zydus Lifesciences Ltd.
System
Details
The airflow visualization study conducted on Line *** Protocol Document CHL-QUA-00315-00, effective June 05, 2021, under dynamic conditions did not demonstrate unidirectional airflow during *** interventions. When the *** to the RABs was *** the airflow (smoke) was observed flowing out of the RABs and into the air intake duct located near the *** line. Air flow on the line was not visible when the *** was *** and thus was not able to be evaluated for unidirectional airflow.
Record Date
FEI Number
Country
Citation Type
Audit Duration (Days)
Investigators
More observations from same firm / company
FDA 483 relevant to Microbiology
Disclaimer: The purpose of this article is to provide helpful information on FDA 483s related to microbiological issues. This article represents 483 information in Tabular form for quick access and easy navigation. Information for this article is derived from “Form FDA 483” (Inspectional Observation). This article is not intended to replace the information provided by the FDA, nor is it guaranteed to be complete, always refer official FDA site for details.