Microbiology FDA 483 (Sentara Enterprises): Smoke studies were inadequately performed under dynamic conditions
Observation
Facility Type
Firm / Company
Sentara Enterprises
System
Details
Specifically, unidirectional airflow was not verified under dynamic conditions representative of your typical production process. Smoke studies conducted in May 2023 in your ISO 5 laminar air flow hoods equipment ID# ***) did not show manipulations or conditions performed *** automated compounding device or repeater pump in use) that would be representative of the dynamic process used in actual production processes.
Record Date
FEI Number
Country
Citation Type
Audit Duration (Days)
Investigators
More observations from same firm / company
FDA 483 relevant to Microbiology
Disclaimer: The purpose of this article is to provide helpful information on FDA 483s related to microbiological issues. This article represents 483 information in Tabular form for quick access and easy navigation. Information for this article is derived from “Form FDA 483” (Inspectional Observation). This article is not intended to replace the information provided by the FDA, nor is it guaranteed to be complete, always refer official FDA site for details.